The C-7100/7200 Series UV/Vis Spectrophotometer: A Seamless Blend of Technology and Design
Innovative Design and Advanced Technology
The C-7100/7200 series UV/Vis spectrophotometer combines cutting-edge technology with a sleek design, making it ideal for laboratories and research facilities. Its modern appearance, ergonomic layout, and smooth contours ensure ease of use without sacrificing performance.
Key features include:
- State-of-the-art microcomputer technology: Enhances electronic control and optimizes optical components for peak performance.
- 1200 lines/mm holographic grating: Ensures precise wavelength dispersion, offering high-quality spectral data with low stray light interference.
- 7-inch color LCD screen: Provides a clear, vibrant display, making data interpretation easy and accessible.
- Silicone buttons: Offer tactile feedback, allowing researchers to navigate functions effortlessly.
The design emphasizes accuracy and stability by minimizing vibrations and maintaining calibration over time. The blend of thoughtful engineering and advanced technology makes the C-7100/7200 a top choice for precise spectrophotometric analysis.
Comprehensive Features for Versatile Testing
The C-7100/7200 series is packed with features that cater to diverse scientific research needs. These spectrophotometers offer:
- Extensive data storage: Helps users efficiently track and retrieve experimental results, ensuring organized data management.
- Multi-format compatibility: Supports various data formats, making results easy to export for reporting and presentations.
- Multiple connectivity options: USB and RS232 ports allow seamless integration with computers and peripherals, enhancing operational convenience.
The spectrophotometers support multiple testing modes:
- Absorbance, transmittance, and energy measurement
- Multi-wavelength testing: Ideal for analyzing complex samples.
- Kinetics measurements: Crucial for studying chemical reaction dynamics over time.
These instruments excel in specialized applications such as DNA and protein quantification, making them essential tools for molecular biology and biochemistry labs.
With easy data output to printers and powerful PC software for in-depth analysis, the C-7100/7200 series ensures researchers can efficiently generate reports and achieve reliable, accurate results in a wide range of laboratory tasks.
Specifications
Model | C-7100 | C-7100S | C-7100A | C-7200 | C-7200S | C-7200A |
Display | 7 inch LCD screen | |||||
Keyboard Control | Silicone Button | |||||
Optical System | Single Beam | Double Beam | ||||
Holographic grating, 1200 lines/mm | ||||||
Spectral Bandwidth | 2nm | 1nm | 0.5,1,2,4nm | 2nm | 1nm | 0.5,1,2,4nm |
Wavelength Range | 190 – 1100nm | |||||
Wavelength Accuracy | ±0.3nm | |||||
Wavelength Repeatability | ≤0.2nm | |||||
Photometric Accuracy | 0.2%T(0-100%T), ±0.002A(0-0.5A), ±0.004A(0.5-1A) | |||||
Photometric Repeatability | ≤0.15%T(0-100%T), 0.001A(0-0.5A), 0.002A(0.5-1A) | |||||
Stray Light | ≤0.03%T@220nm,360nm | |||||
Stability | ±0.002A/h@500nm | |||||
Photometric Range | 0-200%T, -0.3-3.0A, 0-9999C(0-9999F) | |||||
Baseline Flatness | ±0.002A(200-1000nm) | |||||
Noise | 0.0003A@500nm | |||||
Working Mode | T,A,C,E | |||||
Wavelength Setting | Automatic | |||||
Scanning Speed | High, Medium and Low | |||||
Detector | Solid Silicon Photodiode | |||||
Light Source | Tungsten Halogen/Deuterium Lamp | |||||
Data Output | RS232, USB(A), USB(B) | |||||
Processor | Cortex_M3, 120Mhz | |||||
Power Requirements | AC 110-220V 50-60Hz | |||||
Shipping Dimensions & Weight | 770*630*340mm, 27kg | 940*740*510mm
52KG |
770*630*340mm, 27kg | 940*740*510mm
52KG |